A case of exhaustion of implantable cardioverter defibrillator therapies due to inappropriate programming.

ΤίτλοςA case of exhaustion of implantable cardioverter defibrillator therapies due to inappropriate programming.
Publication TypeJournal Article
Year of Publication2011
AuthorsFragakis, N. K., Antoniadis A. P., & Geleris P.
JournalHellenic J Cardiol
Volume52
Issue4
Pagination354-6
Date Published2011 Jul-Aug
ISSN2241-5955
Λέξεις κλειδιάDefibrillators, Implantable, Equipment Failure, Female, Humans, Medical Errors, Middle Aged, Tachycardia, Ventricular, Ventricular Fibrillation
Abstract

Strategically chosen tachycardia detection and therapy options targeting non-sustained ventricular tachycardia (VT) or ventricular fibrillation (VF) can efficiently reduce the morbidity related to implantable cardioverter defibrillator (ICD) therapy. We report a case of a 60-year-old woman with ischemic cardiomyopathy who underwent ICD implantation due to frequent episodes of non-sustained VT. In this case, the inappropriate setting for VF detection in addition to the limited flexibility in device programming for tachycardia confirmation led to the rapid exhaustion of all available shock therapies.