Ultra-low-dose coronary artery calcium scoring using novel scoring thresholds for low tube voltage protocols-a pilot study.

ΤίτλοςUltra-low-dose coronary artery calcium scoring using novel scoring thresholds for low tube voltage protocols-a pilot study.
Publication TypeJournal Article
Year of Publication2018
AuthorsGräni, C., Vontobel J., Benz D. C., Bacanovic S., Giannopoulos A. A., Messerli M., Grossmann M., Gebhard C., Pazhenkottil A. P., Gaemperli O., Kaufmann P. A., & Buechel R. R.
JournalEur Heart J Cardiovasc Imaging
Date Published2018 Feb 08
ISSN2047-2412
Abstract

Aims: To determine if tube-adapted thresholds for coronary artery calcium (CAC) scoring by computed tomography at 80 kilovolt-peak (kVp) tube voltage and 70-kVp yield comparable results to the standard 120-kVp protocol.Methods and results: We prospectively included 103 patients who underwent standard scanning with 120-kVp tube voltage and additional scans with 80 kVp and 70 kVp. Mean body mass index (BMI) was 27.9 ± 5.1 kg/m2. For the lowered tube voltages, we applied novel kVp-adapted thresholds for calculation of CAC scores and compared them with standard 120-kVp scans using intraclass correlation and Bland-Altman (BA) analysis. Furthermore, risk-class (CAC score 0/1-10/11-100/101-400/>400) changes were assessed. Median CAC score from 120-kVp scans was 212 (interquartile range 25-901). Thirteen (12.6%) patients had zero CAC. Using the novel kVp-adapted thresholds, CAC scores derived from 80-kVp scans showed excellent correlation (r = 0.994, P < 0.001) with standard 120-kVp scans with BA limits of agreement of -235 (-39.5%) to 172 (28.9%). Similarly, for 70-kVp scans, correlation was excellent (r = 0.972, P < 0.001) but with broader limits of agreement of -476 (-85.0%) to 270 (48.2%). Only 2 (2.8%) reclassifications were observed for the 80-kVp scans in patients with a BMI <30 kg/m2 (n = 71), and 2 (6.1%) for the 70-kVp scans in patients with a BMI <25 kg/m2 (n = 33). Mean effective radiation dose was 0.60 ± 0.07 millisieverts (mSv), 0.19 ± 0.02 mSv, and 0.12 ± 0.01 mSv for the 120-kVp, 80-kVp, and 70-kVp scans, respectively.Conclusion: The present study suggests that CAC scoring with reduced peak tube voltage is accurate if kVp-adapted thresholds for calculation of CAC scores are applied while offering a substantial further radiation dose reduction.

DOI10.1093/ehjci/jey019
Alternate JournalEur Heart J Cardiovasc Imaging
PubMed ID29432592